The clinical application of a urine albumin:creatinine ratio point-of-care device.

作者信息

Croal B L, Mutch W J, Clark B M, Dickie A, Church J, Noble D, Ross I S

机构信息

Health Services Research Unit, University of Aberdeen, AB515JE, Aberdeen, UK.

出版信息

Clin Chim Acta. 2001 May;307(1-2):15-21. doi: 10.1016/s0009-8981(01)00450-8.

Abstract

INTRODUCTION

Microalbuminuria is an accepted predictive marker for the early detection of renal disease and the identification of patients at high risk of developing complications of diabetes and hypertension. The Bayer Clinitek 50 is a urine chemistry point-of-care analyser for the semi-quantitative measurement of albumin and creatinine and calculation of albumin:creatinine ratio (ACR).

METHOD

Urine samples were obtained from 252 consecutive patients attending a city center diabetic clinic, and from 40 patients on admission to the ICU. Albumin and creatinine measurements were carried out using the Clinitek 50 and by the central laboratory.

RESULTS

The Clinitek 50 results agreed with the central laboratory results in 89% of the diabetic patient samples and 80% of the ICU patient samples. Excluding samples defined as normal by the Clinitek 50 (ACR<3.4 mg/mmol) would have resulted in an 80% reduction in samples sent to the lab for further quantification. The average length of stay in the group of ICU patients with normal ACR was significantly less than for those patients with an abnormal ACR (p<0.005).

CONCLUSIONS

The Clinitek 50 provides useful, immediate clinical information regarding the microalbuminuria status for use in the diabetic clinic setting or as a potential immediate risk management tool in intensive care.