The authors report a case in which the cut end of a Raimondi peritoneal catheter apparently caused intestinal perforation and contamination of the shunt system. The authors advise that if this type of catheter is to be divided the cut should be distal to the end of the sharp wire contained in it, which otherwise will protrude.
作者报告了一例雷蒙迪腹膜导管的断端明显导致肠穿孔和分流系统污染的病例。作者建议,如果要切断这种类型的导管,切口应在其所含尖锐钢丝末端的远端,否则钢丝会突出。