Morley R E, Richter E J, Klaesner J W, Maluf K S, Mueller M J
Department of Electrical Engineering, Washington University, St Louis, MO 63130-4899, USA.
IEEE Trans Biomed Eng. 2001 Jul;48(7):815-20. doi: 10.1109/10.930906.
Patients with diabetes and peripheral neuropathy are susceptible to unnoticed trauma on the foot that can cause skin breakdown. We have designed an electronic system in a shoe that monitors temperature, pressure, and humidity, storing the data in a battery-powered device for later uploading to a host computer for data analysis. The pressure sensors are located at the heel, and under three metatarsal heads. Temperature sensors are located under the medial metatarsal head and under the heel. The humidity sensor is located in the toe of the shoe. Correlations of data from pressure sensors with known values were high (r > 0.85), even after extended use. Although data currently are being collected for descriptive purposes, the design potentially can be used to provide feedback to patients.
患有糖尿病和周围神经病变的患者容易在足部受到未被察觉的创伤,进而导致皮肤破损。我们设计了一种置于鞋内的电子系统,该系统可监测温度、压力和湿度,并将数据存储在一个电池供电的设备中,以便后续上传至主机进行数据分析。压力传感器位于脚跟以及三个跖骨头下方。温度传感器位于内侧跖骨头下方和脚跟下方。湿度传感器位于鞋尖处。即便在长期使用后,压力传感器的数据与已知值之间的相关性依然很高(r > 0.85)。尽管目前收集数据是用于描述性目的,但该设计有可能用于为患者提供反馈。