Leino A, Saarimies J, Grönholm M, Lillsunde P
Department of Clinical Chemistry, University Hospital, Turku, Finland.
Scand J Clin Lab Invest. 2001 Jul;61(4):325-31. doi: 10.1080/00365510152379067.
Eight commercially available on-site drugs-of-abuse testing devices for detecting cannabinoids (THC-COOH), opiates (OPI), cocaine (COC), amphetamines (AMP), metamphetamines (MET) and benzodiazepines (BZO) were evaluated. The used urine specimens suspected of being drug positive were all confirmed by gas chromatographic/mass spectrometry (GC/MS). For AMP and MET, sensitivities varied between 83 and 95% and specificities between 98 and 100%. Correspondingly, sensitivities between 88 and 98% and specificities between 95 and 100% were observed for THC-COOH. For BZO, sensitivities varied between 91 and 97% and specificities between 97 and 100%. Only a few confirmed positive samples were available for OPI and COC, the sensitivities being between 83 and 100% and 100%, respectively. On-site devices did not always find extremely high drug concentrations. False-negative results were found with AMP in particular. Pholcodine, commonly used as medicine, was observed to give false-positive results with most of the devices and was not, however, included in given cross-reactivity tables. It was found that the devices differed markedly with respect to interpretation of test results and to ease of test performance, leading to the suggestion that different criteria for selecting on-site devices for either emergency laboratories in hospitals or for police stations and prisons should be used. Since the overall specificity of any of the devices was not 100% and false positives were identified, we found it important to confirm any positive screening test result.
对8种市售的现场滥用药物检测设备进行了评估,这些设备用于检测大麻素(四氢大麻酚酸,THC-COOH)、阿片类药物(OPI)、可卡因(COC)、苯丙胺(AMP)、甲基苯丙胺(MET)和苯二氮䓬类药物(BZO)。所有怀疑药物呈阳性的尿液标本均通过气相色谱/质谱联用仪(GC/MS)进行了确认。对于AMP和MET,灵敏度在83%至95%之间,特异性在98%至100%之间。相应地,THC-COOH的灵敏度在88%至98%之间,特异性在95%至100%之间。对于BZO,灵敏度在91%至97%之间,特异性在97%至100%之间。OPI和COC仅有少量经确认的阳性样本,其灵敏度分别在83%至100%和100%之间。现场检测设备并非总能检测到极高的药物浓度。尤其在检测AMP时出现了假阴性结果。观察发现,常用作药物的福尔可定在大多数设备检测中会产生假阳性结果,然而它并未包含在给定的交叉反应表中。结果发现,这些设备在检测结果的判读和检测操作的简便程度方面存在显著差异,这表明在为医院急诊实验室、警察局和监狱选择现场检测设备时应采用不同的标准。由于任何一种设备的总体特异性都不是100%,且存在假阳性情况,因此我们认为对任何阳性筛查检测结果进行确认非常重要。