Novel drug-delivery stent: intravascular ultrasound observations from the first human experience with the QP2-eluting polymer stent system.

作者信息

Honda Y, Grube E, de La Fuente L M, Yock P G, Stertzer S H, Fitzgerald P J

机构信息

Center for Research in Cardiovascular Interventions, Stanford University, Stanford, California, USA.

出版信息

Circulation. 2001 Jul 24;104(4):380-3. doi: 10.1161/hc2901.094149.

Abstract

BACKGROUND

The aim of this study was to use serial intravascular ultrasound (IVUS) to evaluate the long-term effect of stent-based 7-hexanoyltaxol (QP2, a taxane analogue) delivery on neointimal tissue growth within the stent and on vessel dimensions at the adjacent reference segments.

METHODS AND RESULTS

Serial IVUS analyses (immediately after intervention and at follow-up at 8.3 months) were performed in 15 native coronary lesions treated with the QuaDS-QP2 stent. IVUS measurements were performed at 8 cross-sections in each target segment (4 cross-sections within the stent and 2 cross-sections in each reference segment). At baseline, no significant plaque protrusion or thrombus was detected in the target segment. Mild incomplete stent apposition and edge dissection were observed in one and two cases, respectively. Percent expansion of the stent (minimum stent area/average reference lumen area) was 96.0+/-21.7%. At follow-up, mean neointimal area within the stent was 1.2+/-1.3 mm(2), and mean cross-sectional narrowing (neointimal area/stent area) was 13.6+/-14.9%. At the vessel segments immediately adjacent to the stent, a significant increase in plaque area (1.9+/-2.6 mm(2), P=0.001) was observed, but vessel area remained unchanged. However, no patients showed clinically significant in-stent or edge restenosis (diameter stenosis >/=50%) during the follow-up period.

CONCLUSIONS

The first human experience with the new drug-delivery stent showed a minimal amount of neointimal proliferation in the stented segment. Late lumen loss at the reference sites adjacent to the stent was acceptable and predominantly due to plaque proliferation.

摘要

背景

本研究旨在使用连续血管内超声(IVUS)评估基于支架的7-己酰紫杉醇(QP2,一种紫杉烷类似物)递送对支架内新生内膜组织生长以及相邻参考节段血管尺寸的长期影响。

方法与结果

对15例接受QuaDS-QP2支架治疗的原发性冠状动脉病变进行了连续IVUS分析(干预后立即及8.3个月随访时)。在每个靶节段的8个横截面进行IVUS测量(支架内4个横截面,每个参考节段2个横截面)。基线时,在靶节段未检测到明显的斑块突出或血栓。分别在1例和2例中观察到轻度不完全支架贴壁和边缘夹层。支架的扩张百分比(最小支架面积/平均参考管腔面积)为96.0±21.7%。随访时,支架内平均新生内膜面积为1.2±1.3 mm²,平均横截面狭窄(新生内膜面积/支架面积)为13.6±14.9%。在紧邻支架的血管节段,观察到斑块面积显著增加(1.9±2.6 mm²,P = 0.001),但血管面积保持不变。然而,在随访期间,没有患者出现临床上显著的支架内或边缘再狭窄(直径狭窄≥50%)。

结论

新型药物洗脱支架的首次人体试验显示,支架置入节段的新生内膜增殖量极少。支架相邻参考部位的晚期管腔丢失是可以接受的,主要是由于斑块增殖。