Buratti S, Cosio M S, Benedetti S, Mannino S
Dipartimento di Scienze e Tecnologie Alimentari e Microbiologiche, Università degli Studi di Milano, Italy.
Analyst. 2001 Aug;126(8):1466-8. doi: 10.1039/b103192a.
In this paper a rapid method based on a FIA (flow injection analysis) system with amperometric detection for the evaluation of ascorbyl 6-palmitate in foods is described. The selectivity of the proposed method is related to the low anodic potential applied to the working glassy carbon electrode (+0.1 V vs. Ag/AgCl) that leaves out interferences from ascorbic acid and phenolic compounds. By flow injection analysis, under optimised conditions, the calibration curve was linear in the range 0-20 mg l(-1) and the detection limit was 0.2 mg l(-1).
本文描述了一种基于流动注射分析(FIA)系统并采用安培检测法来评估食品中抗坏血酸6-棕榈酸酯的快速方法。所提出方法的选择性与施加于工作玻碳电极上的低阳极电位(相对于Ag/AgCl为+0.1 V)有关,该电位排除了来自抗坏血酸和酚类化合物的干扰。通过流动注射分析,在优化条件下,校准曲线在0 - 20 mg l⁻¹范围内呈线性,检测限为0.2 mg l⁻¹。