Laessig R H, Schwartz T H, Paskey T A, Indriksons A I
Am J Clin Pathol. 1975 Aug;64(2):225-35. doi: 10.1093/ajcp/64.2.225.
A blind quality control system using a special type of commercial reference serum is described. As packaged, the product contains six pools of analyzed reference sera with vials identified by coded numbers. The test constituents in the pools cover the range of multichannel analyzers in a random manner, i.e., each pool contains constituents in the low, normal and elevated ranges. As specific pools are not identified, and the analyst is unaware of the expected values for each constituent, results collected for quality control purposes are unbiased, i.e., free of conscious or unconscious editing. Data analysis in the proposed system is by least-squares technics. During an eight-week period, performance was studied using the standard error of the estimate to evaluate precision and the slope of the regression line to evaluate accuracy on the SMA and Mark X instruments. As an empirical comparison, the data are expressed as relative errors and precision and accuracy of analyzer performance evaluated as mean and standard deviation. The proposed regression analysis-based technic has the advantage of simultaneously evaluating linearity, as well as checking accuracy and precision over the total dynamic range of the instrument by use of a series of related serum pools.
描述了一种使用特殊类型商业参考血清的盲法质量控制系统。该产品包装时包含六个分析参考血清池,小瓶用编码数字标识。池中测试成分以随机方式覆盖多通道分析仪的范围,即每个池都包含低、正常和升高范围内的成分。由于未识别特定池,且分析人员不知道每种成分的预期值,因此为质量控制目的收集的结果是无偏的,即没有有意识或无意识的编辑。所提出系统中的数据分析采用最小二乘法。在八周期间,使用估计标准误差评估精密度,使用回归线斜率评估SMA和Mark X仪器的准确度,对性能进行了研究。作为经验比较,数据表示为相对误差,分析仪性能的精密度和准确度以平均值和标准差评估。所提出的基于回归分析的技术具有同时评估线性的优点,并且通过使用一系列相关血清池在仪器的整个动态范围内检查准确度和精密度。