Suppr超能文献

Dimensions of stabident intraosseous perforators and needles.

作者信息

Ramlee R A, Whitworth J

机构信息

Department of Restorative Dentistry, University of Newcastle upon Tyne, Newcastle, UK.

出版信息

J Endod. 2001 Sep;27(9):581-3. doi: 10.1097/00004770-200109000-00007.

Abstract

Problems can be encountered inserting intraosseous injection needles through perforation sites. This in vitro study examined the variability and size compatibility of Stabident intraosseous injection components. The diameters of 40 needles and perforators from a single Stabident kit were measured in triplicate with a toolmakers microscope. One-way ANOVA revealed that mean needle diameter (0.411 mm) was significantly narrower than mean perforator diameter (0.427 mm) (p < 0.001). A frequency distribution plot revealed that needle diameter followed a normal distribution, indicating tight quality control during manufacture. The diameter of perforators was haphazardly distributed, with a clustering of 15% at the lower limit of the size range. However on no occasion was the diameter of a perforator smaller than that of an injection needle. We conclude that components of the Stabident intraosseous anaesthetic system are size-compatible, but there is greater and more haphazard variability in the diameter of perforators than injection needles.