Suppr超能文献

用于白内障手术中减少超声时间和能量的喇叭状超声乳化头

Flared phacoemulsification tips to decrease ultrasound time and energy in cataract surgery.

作者信息

McNeill J I

机构信息

Department of Ophthalmology, Loma Linda University School of Medicine, Loma Linda, California, USA.

出版信息

J Cataract Refract Surg. 2001 Sep;27(9):1433-6. doi: 10.1016/s0886-3350(01)00789-1.

Abstract

PURPOSE

To determine the effect on ultrasound (US) time and energy of using flared phacoemulsification tips in cataract surgery.

SETTING

Private referral practice of Pacific Cataract & Laser Institute, Chehalis, Washington, USA.

METHODS

One surgeon using the Allergan Medical Optics Diplomate phacoemulsification machine performed cataract surgery by a single divide-and-conquer technique. A prospective nonmasked consecutive study group (n = 262) was selected from patients operated on with phaco tips flared by 25%. A consecutive control group (n = 440) was selected from patients having surgery using the same technique except for the use of straight phaco tips. The study group had cataract surgery over a 5 week period and the control group, over an earlier 10 week period. Two weeks separated the groups. Cases complicated by vitreous loss were eliminated. The 2 groups were compared by age, sex, total phaco time (TPT, the total time ultrasound energy was applied), and equivalent phaco time (EPT, the product of the phaco time elapsed and the phaco energy).

RESULTS

One patient in each group was eliminated for vitreous loss during surgery. There were no significant differences in age, sex, or eye between the 2 groups. The study group had a mean TPT of 29.2 seconds +/- 11.7 (SD) (range 0 to 88 seconds) and a mean EPT of 16.4 +/- 8.6 seconds (range 0 to 67 seconds). The control group had a mean TPT of 34.4 +/- 16.1 seconds (range 0 to 130 seconds) (P <.001) and a mean EPT of 20.0 +/- 12.8 seconds (range 0 to 107 seconds) (P <.001).

CONCLUSION

Flared ultrasonic phacoemulsification tips significantly reduced the US time and energy used in cataract surgery.

摘要

目的

确定在白内障手术中使用扩口式超声乳化针头对超声(US)时间和能量的影响。

设置

美国华盛顿州切哈里斯市太平洋白内障与激光研究所的私人转诊诊所。

方法

一名外科医生使用爱尔康医疗光学公司的Diplomate超声乳化仪,通过单一的分而治之技术进行白内障手术。从使用扩口25%的超声乳化针头进行手术的患者中选取前瞻性非盲连续研究组(n = 262)。从使用相同技术但使用直型超声乳化针头进行手术的患者中选取连续对照组(n = 440)。研究组在5周内进行白内障手术,对照组在更早的10周内进行手术。两组间隔两周。排除玻璃体脱出并发症的病例。比较两组患者的年龄、性别、总超声乳化时间(TPT,即应用超声能量的总时间)和等效超声乳化时间(EPT,即超声乳化经过时间与超声乳化能量的乘积)。

结果

每组各有1例患者在手术中因玻璃体脱出被排除。两组患者在年龄、性别或患眼方面无显著差异。研究组的平均TPT为29.2秒±11.7(标准差)(范围0至88秒),平均EPT为16.4±8.6秒(范围0至67秒)。对照组的平均TPT为34.4±16.1秒(范围0至130秒)(P <.001),平均EPT为20.0±12.8秒(范围0至107秒)(P <.001)。

结论

扩口式超声乳化针头显著减少了白内障手术中使用的超声时间和能量。