Berrocal T, Gayá F, Arjonilla A, Lonergan G J
Department of Pediatric Radiology, University Hospital La Paz, Paseo de la Castellana 261, 28046 Madrid, Spain.
Radiology. 2001 Nov;221(2):359-65. doi: 10.1148/radiol.2212001807.
To evaluate the usefulness of echo-enhanced cystosonography compared with voiding cystourethrography (VCUG) for detecting and grading vesicoureteral reflux (VUR).
Two hundred sixteen pediatric patients underwent cystosonography enhanced with SH U 508A, a galactose-based echo-enhancing agent. Sonograms of the kidneys and bladder were obtained before filling, during bladder filling, and during voiding. This examination was followed immediately with VCUG. Each kidney or portion of a kidney with its own complete collecting system was considered separately, for a total of 440 kidney units.
VUR was detected in 123 kidney units at cystosonography and in 104 at VCUG. In 401 kidney units, there was concordance between results at cystosonography and at VCUG regarding the presence or absence of VUR. Ninety-four kidney units showed VUR with both methods. Seventy-seven of the 94 refluxing units were depicted with the same grade of VUR with both modalities, and in 17 the VUR grade was greater at cystosonography than at VCUG. Twenty-nine of the 94 units showed VUR at only cystosonography, and 10 units at only VCUG. The McNemar test showed that cystosonography depicted a significantly (P = .003) higher number of units with VUR. By patient, VUR was depicted with both studies in 67 and with only one study in 25. VUR was seen at only cystosonography in 16 patients and at only VCUG in nine. The McNemar test for patients showed no significant difference between the two tests in detection of VUR.
Cystosonography with SH U 508A appears comparable to VCUG in the depiction of VUR.
评估超声造影增强的膀胱超声检查与排尿性膀胱尿道造影(VCUG)在检测和分级膀胱输尿管反流(VUR)方面的实用性。
216例儿科患者接受了用基于半乳糖的超声造影剂SH U 508A增强的膀胱超声检查。在膀胱充盈前、充盈期间及排尿期间获取肾脏和膀胱的超声图像。此检查后立即进行VCUG。每个具有完整集合系统的肾脏或肾脏部分被分别考虑,共计440个肾单位。
膀胱超声检查发现123个肾单位存在VUR,VCUG发现104个。在401个肾单位中,膀胱超声检查和VCUG在VUR的有无结果上具有一致性。两种方法均显示94个肾单位存在VUR。94个反流单位中的77个在两种检查方式下VUR分级相同,17个在膀胱超声检查时VUR分级高于VCUG。94个单位中,29个仅在膀胱超声检查时显示VUR,10个仅在VCUG时显示。McNemar检验显示,膀胱超声检查发现的VUR单位数量显著更多(P = 0.003)。按患者计算,两种检查均显示VUR的有67例,仅一种检查显示的有25例。仅在膀胱超声检查时发现VUR的有16例患者,仅在VCUG时发现的有9例。患者的McNemar检验显示,两种检查在检测VUR方面无显著差异。
使用SH U 508A的膀胱超声检查在显示VUR方面似乎与VCUG相当。