Boudet G, Chaumoux A
Laboratoire Performance Motrice, UFR STAPS, Université Blaise Pasal, Aubière, Cedex, France.
J Sports Med Phys Fitness. 2001 Dec;41(4):546-53.
The extended use of heart rate monitors to non sportive people, and older ones increases the risk to be facing heart rate troubles. Questions exist upon the ability of this devices to detect such abnormalities. The purpose of the investigation was to evaluate the accuracy of two third generation heart rate monitors, Accurex Plus and Vantage NV and to compared these data with those of an older one, PE 4000. Then we investigate responses of this monitors to abnormal heart rate.
The three heart rate monitors were tested in the laboratory, wired in parallel to an ECG simulator under two modes: a normal heart rate program (7 stable heart rate stages with two marked transitions) and an automatic arrhythmia one.
1st program: Values generated by the ECG simulator were compared to the values produced by the three devices. Particular attention was paid to the transition phases (heart rate drop and heart rate fall). Results of heart rate monitors accuracy were expressed at the exact value (devices value=simulator value) and at approximated value of +/-3 beats x min(-1) (devices value=simulator value +/-3 beats x min(-1)); 2nd program: Abnormal rhythms were analysed using an ambulatory ECG recorder (Synésis) as control data, and compared to the data from the three devices.
Correlation between simulator generated true values and HRMs read values were calculated.
In the 1st normal heart rate program, with an accuracy of exact value, Accurex Plus and Vantage NV, were more accurate than the PE 4000 (94% and 89% of values respectively versus 33% for the PE4000). At +/-3 beats x min(-1), the three devices gave good results: over 98% of total values. In transition phases: the three devices showed a smoothing effect, which was stronger in decelerating heart rate than in accelerating heart rate. In the 2nd program: isolated heart rhythm troubles (missing beat, pause <4 sec, supra-ventricular ectopic activity, and ventricular ectopic activity) were either not detected or were rejected. Signal changes such as changes in QRS (ventricular bigeminy) may cause detection losses, and so modify calculated heart rate. The HR signal rising/dropping slopes caused by sudden heart rhythm disorders, were notably attenuated, with a consequent loss of HR accuracy.
In laboratory stable heart rate conditions, third generation HRMs are more accurate than earlier ones. Heart rate monitors are less accurate in transient phases and have not been improved in that domain. The three HRMs ignore isolated heart rate troubles. As expected, the usefulness of HRMs in detecting HR disorders is limited. Nevertheless, these devices have some value with serious troubles, like pause, bradycardia or tachycardia, lasting longer than 4 sec, and especially if they coincide with functional symptoms. The use of such even more accurate devices may consequently be recommended to healthy public for which they were built.
心率监测器在非运动人群及老年人中的广泛使用增加了面临心率问题的风险。对于此类设备检测此类异常的能力存在疑问。本研究的目的是评估两款第三代心率监测器Accurex Plus和Vantage NV的准确性,并将这些数据与一款较旧的产品PE 4000的数据进行比较。然后我们研究了这些监测器对异常心率的反应。
三款心率监测器在实验室中进行测试,与一台心电图模拟器并联连接,测试模式有两种:正常心率程序(7个稳定心率阶段,有两个明显的转变)和自动心律失常模式。
第一个程序:将心电图模拟器生成的值与三款设备产生的值进行比较。特别关注转变阶段(心率下降和心率上升)。心率监测器准确性的结果以精确值(设备值 = 模拟器值)和近似值±3次/分钟(设备值 = 模拟器值±3次/分钟)表示;第二个程序:使用动态心电图记录仪(Synésis)作为对照数据分析异常心律,并与三款设备的数据进行比较。
计算模拟器生成的真实值与心率监测器读取值之间的相关性。
在第一个正常心率程序中,以精确值的准确性来看,Accurex Plus和Vantage NV比PE 4000更准确(分别为94%和89%的值,而PE4000为33%)。在±3次/分钟时,三款设备都有良好的结果:超过98%的总值。在转变阶段:三款设备都显示出平滑效应,心率减速时比心率加速时更强。在第二个程序中:孤立的心律问题(漏搏、<4秒的停搏、室上性异位活动和室性异位活动)要么未被检测到,要么被排除。信号变化如QRS波变化(室性二联律)可能导致检测失败,从而改变计算出的心率。由突然的心律紊乱引起的心率信号上升/下降斜率明显减弱,导致心率准确性丧失。
在实验室稳定心率条件下,第三代心率监测器比早期的更准确。心率监测器在瞬态阶段准确性较低,且在该领域未得到改进。三款心率监测器都忽略孤立的心率问题。正如预期的那样,心率监测器在检测心率紊乱方面的作用有限。然而,这些设备对于严重问题,如持续超过4秒的停搏、心动过缓或心动过速,尤其是与功能症状同时出现时,具有一定价值。因此,对于为此类设备设计的健康人群,可能推荐使用更准确的设备。