Spuentrup Elmar, Manning Warren J, Botnar René M, Kissinger Kraig V, Stuber Matthias
Department of Medicine, Cardiovascular Division, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts 02215, USA.
Magn Reson Med. 2002 Jan;47(1):196-201. doi: 10.1002/mrm.10026.
The purpose of this study was to investigate the impact of navigator timing on image quality in navigator-gated and real-time motion-corrected, free-breathing, three-dimensional (3D) coronary MR angiography (MRA) with submillimeter spatial image resolution. Both phantom and in vivo investigations were performed. 3D coronary MRA with real-time navigator technology was applied using variable navigator time delays (time delay between the navigator and imaging sequences) and varying spatial resolutions. Quantitative objective and subjective image quality parameters were assessed. For high-resolution imaging, reduced image quality was found as a function of increasing navigator time delay. Lower spatial resolution coronary MRA showed only minor sensitivity to navigator timing. These findings were consistent among volunteers and phantom experiments. In conclusion, for submillimeter navigator-gated and real-time motion-corrected 3D coronary MRA, shortening the time delay between the navigator and the imaging portion of the sequence becomes increasingly important for improved spatial resolution.
本研究的目的是探讨在具有亚毫米空间图像分辨率的导航门控和实时运动校正、自由呼吸三维(3D)冠状动脉磁共振血管造影(MRA)中,导航器时间对图像质量的影响。同时进行了体模和体内研究。使用可变的导航器时间延迟(导航器与成像序列之间的时间延迟)和不同的空间分辨率,应用具有实时导航技术的3D冠状动脉MRA。评估了定量的客观和主观图像质量参数。对于高分辨率成像,发现图像质量随着导航器时间延迟的增加而降低。较低空间分辨率的冠状动脉MRA对导航器时间仅表现出较小的敏感性。这些发现在志愿者和体模实验中是一致的。总之,对于亚毫米导航门控和实时运动校正的3D冠状动脉MRA,缩短导航器与序列成像部分之间的时间延迟对于提高空间分辨率变得越来越重要。