Relationship between phantom failure rates and radiation dose in mammography accreditation.

作者信息

Haus A G, Yaffe M J, Feig S A, Hendrick R E, Butler P A, Wilcox P A, Bansal S

机构信息

Sunnybrook & Womens College Health Sciences Center, Toronto, Canada.

出版信息

Med Phys. 2001 Nov;28(11):2297-301. doi: 10.1118/1.1408283.

Abstract

The American College of Radiology Mammography Accreditation Program (ACR MAP) reviews both clinical mammograms and a phantom image to assess clinical and technical quality from each mammography unit. The phantom contains details representing fibers (speculations), speck groups (microcalcifications), and masses. The depiction of these structures by the mammographic system is scored by medical physicists. The phantom image is taken using the facility's exposure technique for a 4.2-cm thick breast of average composition. The mean glandular dose (MGD) is determined from a set of thermoluminescent dosimeters placed on top of the chest wall edge of the phantom. Phantom scores and MGD data collected from 1993 to 1999 based on 31 535 unit evaluations are presented in this paper. The relationship between the failure rate for phantom image quality and MGD has been analyzed. While over all doses the phantom failure rate was 11%, for doses of 0.26 to 0.50 mGy the failure rate was 43%. The phantom failure rate fell continuously to about 6% for MGDs in the range of 1.51-2.0 mGy. With further increases in dose, failure rates began to rise. Factors that may account for these results are presented and discussed.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验