Lawrence Ben G, Crozier Stuart, Yau Desmond D, Doddrell David M
Centre for Magnetic Resonance, University of Queensland, St Lucia Brisbane, Australia.
IEEE Trans Biomed Eng. 2002 Jan;49(1):64-71. doi: 10.1109/10.972841.
An inverse methodology is described to assist in the design of radio-frequency (RF) coils for magnetic resonance imaging (MRI) applications. The time-harmonic electromagnetic Green's functions are used to calculate current on the coil and shield cylinders that will generate a specified internal magnetic field. Stream function techniques and the method of moments are then used to implement this theoretical current density into an RF coil. A novel asymmetric coil operating for a 4.5 T MRI machine was designed and constructed using this methodology and the results are presented.