Cross-stimulation during lead impedance monitoring.

作者信息

Beaufort-Krol G C, Nagelkerke D, Van Gelder B M

机构信息

Beatrix Children's Hospital Groningen, Division of Pediatric Cardiology University Hospital, The Netherlands.

出版信息

Pacing Clin Electrophysiol. 2001 Nov;24(11):1696-8. doi: 10.1046/j.1460-9592.2001.01696.x.

Abstract

At the age of 4 years, a total cavopulmonary connection was performed in a boy with a complex congenital heart defect. On addition, a DDDR pacemaker was implanted for sick sinus syndrome. Atrial and ventricular leads were epicardially placed at the left atrium and left ventricle. At the age of 10 years, a new epicardial ventricular lead was placed because of malfunction of the existing lead. At the same operation the pulse generator was replaced by a Medtronic Kappa DR 731. After replacement, the boy experienced episodes of phrenic nerve stimulation associated with feelings of discomfort. Holter recordings revealed ventricular stimulation from the atrial stimulus for 2 consecutive beats. This phenomenon repeated exactly every 3 hours and was caused by the automatic lead impedance measurement that used a 5-V, 1-ms stimulus output.

摘要

一名患有复杂先天性心脏缺陷的男孩在4岁时接受了全腔静脉肺动脉连接术。此外,因病态窦房结综合征植入了DDDR起搏器。心房和心室电极经心外膜置于左心房和左心室。10岁时,由于现有电极发生故障,植入了一根新的心外膜心室电极。在同一次手术中,脉冲发生器被美敦力Kappa DR 731取代。更换后,该男孩出现与不适感相关的膈神经刺激发作。动态心电图记录显示,心房刺激连续2次引起心室刺激。这种现象每3小时精确重复一次,是由使用5V、1ms刺激输出的自动电极阻抗测量引起的。