Dynamic changes of cerebrospinal fluid in upright and recumbent shunted experimental animals.

作者信息

Yamada S, Ducker T B, Perot P L

出版信息

Childs Brain. 1975;1(2-3):187-92. doi: 10.1159/000119566.

Abstract

The effect of positional change on intracranial CSF pressure, valve flow rates and atrial pressures were studied in 40 experimental dogs. The flow rate with a 30 to 60 mm H2O Holter valve changed from the recumbent to the head-up position: during the first minute it increased 2 1/2 times and then progressively decreased to become stabilized 20-30 min later at a value slightly lower than the head-up value. At the same time pressure dropped rapidly to -23 mm H2O and then slowly, during 30 min to -120 mm H2O. In dogs without a shunt the pressure drop was initially similar to that of shunted dogs but recovered to almost the initial positive pressure in 30 min. The flow rate from a reservoir to the atrium through a shunt system when changing to the head-up position showed a siphon effect. Atrial pressure does not change with positioning. The findings were similar with cisternal-peritoneal shunts. Flow rate, therefore, increase in the head-up position. Negative intracranial CSF pressure was due to a siphon effect through the shunt system.