Gong Bolin, Wang Yan
Department of Chemistry, Ningxia University, Yinchuan, China.
Anal Bioanal Chem. 2002 Feb;372(4):597-600. doi: 10.1007/s00216-001-1207-5. Epub 2001 Dec 21.
A new polyacrylacylaminothiourea chelating fiber was synthesized simply and rapidly from nitrilon (an acrylonitrile-based synthetic fiber) and aminothiourea. This fiber was used for the pre-concentration and separation of traces of AuIII, PtIV, PdIV, and IrIV ions from aqueous samples. Factors effecting and parameters for the adsorption of these ions onto the resin, such as acidity, rate, reuse, capacity, and interference of other ions, were investigated using inductively-coupled plasma atomic emission spectrometry (ICP-AES). Conditions of desorption of these ions from the resin were similarly investigated. The enrichment factor was 100. The relative standard deviation for the determination of AuIII, PtIV, PdIV, and IrIV were in the range of 0.7-3.0% for 20.0 ng/mL of each. The recoveries of a standard from real solution samples were between 96% and 100%. The concentrations of analyzed ions in a digested metal powder sample processed with the method proposed were in good agreement with the certified values.
一种新型的聚丙烯酰基氨基硫脲螯合纤维由腈纶(一种基于丙烯腈的合成纤维)和氨基硫脲简单快速地合成。该纤维用于从水样中预富集和分离痕量的AuIII、PtIV、PdIV和IrIV离子。使用电感耦合等离子体原子发射光谱法(ICP-AES)研究了影响这些离子吸附到树脂上的因素以及吸附参数,如酸度、速率、重复使用性、容量和其他离子的干扰。同样研究了这些离子从树脂上解吸的条件。富集因子为100。对于每种20.0 ng/mL的AuIII、PtIV、PdIV和IrIV,测定的相对标准偏差在0.7-3.0%范围内。实际溶液样品中标准物的回收率在96%至100%之间。用所提出的方法处理的消化金属粉末样品中分析离子的浓度与认证值高度一致。