Nusynowitz M L, Benedetto A R
J Nucl Med. 1975 Dec;16(12):1200-3.
The modulation transfer function (MTF) has long been accepted as an index of scintillation camera system resolution, but the technique necessary for the derivation of the MTF has discouraged its widespread acceptance because of its tedious and time-consuming nature. We have shown that the generation of LSFs for input into computer programs for MTF generation can be accomplished very easily and in short periods of time by using a simplified method. Establishment of the validity of this simplified procedure should contribute to more widespread utilization of the MTF in routine quality control programs as a check for crystal deterioration, electronic circuitry degradation, and/or collimator damage.
调制传递函数(MTF)长期以来一直被视为闪烁相机系统分辨率的指标,但由于其推导过程繁琐且耗时,用于推导MTF的技术阻碍了它的广泛应用。我们已经表明,通过使用一种简化方法,可以非常轻松且在短时间内生成用于输入到MTF生成计算机程序中的线扩散函数(LSF)。确立这种简化程序的有效性应有助于MTF在常规质量控制程序中更广泛地用于检查晶体退化、电子电路性能下降和/或准直器损坏。