Murphy P H
J Nucl Med. 1987 Jul;28(7):1221-7.
The measurement of several performance parameters of a scintillation camera at the time of installation and thereafter at regular intervals is necessary to ensure that the camera is operating within specifications and to detect changes over time that can initiate a request for service. There are varied opinions on what constitutes a satisfactory acceptance test and quality control program. Few individuals would disagree that either an acceptance test or routine quality control measurements are necessary, yet agreement on the contents and frequencies of these tests is lacking. This paper discusses the performance parameters of scintillation cameras that are usually measured for planar imaging and also for SPECT with rotating cameras.
在安装时以及之后定期测量闪烁相机的几个性能参数,对于确保相机在规格范围内运行以及检测随时间的变化从而引发维修请求是必要的。对于什么构成令人满意的验收测试和质量控制程序存在不同意见。很少有人会不同意验收测试或常规质量控制测量是必要的,然而对于这些测试的内容和频率缺乏共识。本文讨论了通常用于平面成像以及旋转相机SPECT的闪烁相机的性能参数。