Suppr超能文献

体内植入式听觉装置(Ineraid)患者的电听觉诱发电位记录——与心理物理学测量的相关性及对患者适配的潜在意义

EAP recordings in ineraid patients--correlations with psychophysical measures and possible implications for patient fitting.

作者信息

Zimmerling Martin J, Hochmair Erwin S

机构信息

Institute of Applied Physics, University of Innsbruck, Austria.

出版信息

Ear Hear. 2002 Apr;23(2):81-91. doi: 10.1097/00003446-200204000-00001.

Abstract

OBJECTIVE

Objective measurements can be helpful for cochlear implant fitting of difficult populations, as for example very young children. One method, the recording of the electrically evoked compound action potential (EAP), measures the nerve recruitment in the cochlea in response to stimulation through the implant. For coding strategies implemented at a moderate stimulation rate of 250 pps per channel, useful correlations between EAP data and psychophysical data have been already found. With new systems running at higher rates, it is important to check these correlations again.

DESIGN

This study investigates the correlations between psychophysical data and EAP measures calculated from EAP amplitude growth functions. EAP data were recorded in 12 Ineraid subjects. Additionally, behavioral thresholds (THR) and maximum acceptable loudness levels (MAL) were determined for stimulation rates of 80 pps and 2,020 pps for each electrode.

RESULTS

Useful correlations between EAP data and psychophysical data were found at the low stimulation rate (80 pps). However, at the higher stimulation rate (2,020 pps) correlations were not significant. They were improved substantially, however, by introducing a factor that corrected for disparities due to temporal integration. Incorporation of this factor, which controls for the influence of the stimulation rate on the threshold, improved the correlations between EAP measures recorded at 80 pps and psychophysical MALs measured at 2,020 pps to better than r = 0.70.

CONCLUSIONS

EAP data as such can only be used to predict behavioral THRs or MCLs at low stimulation rates. To cope with temporal integration effects at higher stimulation rates, EAP data must be rate corrected. The introduction of a threshold-rate-factor is a promising way to achieve that goal. Further investigations need to be performed.

摘要

目的

客观测量有助于为诸如幼儿等困难群体进行人工耳蜗植入调试。一种方法是记录电诱发复合动作电位(EAP),它可测量耳蜗中神经对通过植入物的刺激的募集情况。对于以每通道250次/秒的中等刺激速率实施的编码策略,已经发现EAP数据与心理物理学数据之间存在有用的相关性。随着新系统以更高的速率运行,再次检查这些相关性很重要。

设计

本研究调查了心理物理学数据与根据EAP振幅增长函数计算出的EAP测量值之间的相关性。在12名使用Ineraid系统的受试者中记录了EAP数据。此外,还确定了每个电极在80次/秒和2020次/秒刺激速率下的行为阈值(THR)和最大可接受响度水平(MAL)。

结果

在低刺激速率(80次/秒)下发现EAP数据与心理物理学数据之间存在有用的相关性。然而,在较高刺激速率(2020次/秒)下,相关性不显著。但是,通过引入一个校正时间整合差异的因子,相关性得到了显著改善。纳入这个控制刺激速率对阈值影响的因子后,在80次/秒记录的EAP测量值与在2020次/秒测量的心理物理学MAL之间的相关性提高到了r>0.70。

结论

仅EAP数据本身只能用于预测低刺激速率下的行为THR或MCL。为了应对较高刺激速率下的时间整合效应,EAP数据必须进行速率校正。引入阈值 - 速率因子是实现该目标的一种有前途的方法。还需要进行进一步的研究。