[HEPFIEx模拟器,一种用于测定股骨头假体与软骨之间摩擦力的装置]

[The HEPFIEx simulator, a device for determination of friction between femur head prosthesis and cartilage].

作者信息

Mehler D, Müller L P, Hely H, Degreif J, Rommens P M

机构信息

Klinik und Poliklinik für Unfallchirurgie, Universitätsklinikum, Johannes Gutenberg-Universität Mainz.

出版信息

Biomed Tech (Berl). 2002 Mar;47(3):63-6. doi: 10.1515/bmte.2002.47.3.63.

Abstract

We describe a device designed to investigate friction between various femoral head prostheses and human acetabula. It enables not only the determination of friction and the relevance of the play between the femoral head and acetabulum, but also the evaluation of the kinematic behaviour of bipolar prostheses. In the simulator, the various femoral head prostheses are placed on a special cone and tested against a human cadaveric acetabulum. The swiveling range of the device is uniaxial, and the swiveling angle is +/- 35 degrees. The maximum force produced pneumatically is 5kN. Testing of the simulator with a TEP was successful and friction-coefficients of < 0.1 were measured, as are reported in the literature.

摘要

我们描述了一种用于研究各种股骨头假体与人体髋臼之间摩擦力的装置。它不仅能够测定摩擦力以及股骨头与髋臼之间间隙的相关性,还能评估双极假体的运动学行为。在模拟器中,将各种股骨头假体放置在一个特殊的圆锥体上,并针对人体尸体髋臼进行测试。该装置的旋转范围是单轴的,旋转角度为±35度。气动产生的最大力为5kN。用一枚全聚乙烯(TEP)对模拟器进行测试获得成功,并测量到摩擦系数<0.1,正如文献中所报道的那样。