Suppr超能文献

Kinetic flow-injection determination of hydrogen peroxide by use of iron(III)-catalyzed coloration and its application to the determination of biological substances.

作者信息

Maeuchihara N, Nakano S, Kawashima T

机构信息

Department of Chemistry, University of Tsukuba, Japan.

出版信息

Anal Sci. 2001 Feb;17(2):255-8. doi: 10.2116/analsci.17.255.

Abstract

A kinetic flow-injection (FI) method is described for the determination of hydrogen peroxide. This method is based on an iron(III)-catalyzed oxidative coupling of 4-aminoantipyrine with N,N-dimethylaniline by hydrogen peroxide. By measuring the change in the absorbance of the dye formed at 560 nm, 1 x 10(-6) - 6 x 10(-4) M hydrogen peroxide could be determined with a sampling rate of 15 h(-1). The relative standard deviation (n = 30) was 0.8% for 5 x 10(-5) M hydrogen peroxide. There was little interference of the co-existing ions and compounds. After introducing some immobilized enzyme reactors to the FI system, the proposed method allowed the determination of glucose and uric acid ranging from 1 x 10(-6) to 6 x 10(-4) M with relative standard deviations of below 2%. The applicability of the method was demonstrated by determining these substances in serum samples.