Abstract

The AtLast Blood Glucose System was developed to eliminate painful finger punctures by testing samples from less sensitive body sites. Studies were designed to evaluate subjects under various glycemic states and to compare glucose concentrations in samples obtained from the forearm and palm to those found in the finger. The palm is more similar to the finger in blood flow rates, while the forearm is less perfused. AtLast measurements of forearm samples from subjects at a minimum of 2 h postprandial produced good regression statistics when compared to reference finger results with a correlation coefficient of 0.978. Agreement was quite weakened in subjects undergoing extreme glucose loads, resulting in a slope of 0.85, an intercept of 30.82, and a correlation coefficient of 0.938. Time course data indicate that forearm glucose levels often appear lower than finger concentrations when glucose is rapidly rising and somewhat higher when glucose is quickly falling. On the other hand, glucose levels of palm samples produced a close correlation to finger concentrations under all glycemic states. In participants undergoing rapid glucose change, the linear correlation resulted in a slope of 1.01, an intercept of 2.46, and a high correlation coefficient of 0.983, with steady tracking of palm and finger concentrations throughout the study. The palm of the hand was also identified as a comfortable testing site that equals the forearm in affording painless testing. Glucose monitoring of blood samples from the forearm is suitable when expecting steady state glycemic conditions. At times of rapid glucose change, the palm of the hand offers a painless sampling site that compares well with the finger.

摘要

AtLast血糖系统的研发目的是通过检测来自不太敏感身体部位的样本,消除痛苦的手指穿刺。研究旨在评估处于各种血糖状态下的受试者,并比较从前臂和手掌获取的样本中的葡萄糖浓度与手指样本中的葡萄糖浓度。手掌在血流速度方面与手指更为相似,而前臂的灌注较少。与参考手指检测结果相比,餐后至少2小时的受试者前臂样本的AtLast测量产生了良好的回归统计数据,相关系数为0.978。在经历极端葡萄糖负荷的受试者中,一致性明显减弱,斜率为0.85,截距为30.82,相关系数为0.938。时间进程数据表明,当葡萄糖快速上升时,前臂葡萄糖水平通常低于手指浓度,而当葡萄糖快速下降时,前臂葡萄糖水平则略高。另一方面,在所有血糖状态下,手掌样本的葡萄糖水平与手指浓度密切相关。在经历快速葡萄糖变化的参与者中,线性相关性导致斜率为1.01,截距为2.46,相关系数高达0.983,在整个研究过程中手掌和手指浓度持续追踪良好。手掌也被确定为一个舒适的检测部位,在提供无痛检测方面与前臂相当。在预期血糖处于稳定状态时,适合监测前臂血样中的葡萄糖。在葡萄糖快速变化时,手掌提供了一个与手指相比无痛的采样部位。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验