Selectseed 125I 组织间近距离治疗籽源的热释光剂量测定法

Thermoluminescent dosimetry of the selectseed 125I interstitial brachytherapy seed.

作者信息

Anagnostopoulos G, Baltas D, Karaiskos P, Sandilos P, Papagiannis P, Sakelliou L

机构信息

Department of Medical Physics and Engineering, Strahlenklinik, Klinikum Offenbach, Germany.

出版信息

Med Phys. 2002 May;29(5):709-16. doi: 10.1118/1.1469631.

Abstract

This work presents experimental dosimetry results for the new selectSeed 125I prostate seed design for use with the seedSelectron afterloading device, in accordance with the AAPM advisory that all new low energy interstitial brachytherapy seeds should undergo one Monte Carlo (MC) and at least one experimental dosimetry characterization. TLD dosimetry was performed using 120 cylindrical LiF TLD type-100 rods calibrated using a 6 MV photon beam. They were irradiated in solid water phantoms for the experimental determination of the seed dose rate constant, radial dose functions and anisotropy functions. MC simulations were performed for the determination of the TLDs relative energy response that was found position independent and equal to 1.40+/-0.03, and for the calculation of the ratio of dose in liquid water to dose in solid water that was found to be well described by Dliquidwater/Dsolidwater= 1.013*r+0.030 presenting only a minor dependence on polar angle. The selectSeed dose rate constant in liquid water was found equal to 0.938+/-0.065 cGy h(-1) U(-1), which agrees within experimental uncertainties with corresponding MC results of lambdaselect Seed=0.954+/-0.005 cGy h(-1) U(-1). The experimental radial dose and anisotropy function results were also found in good agreement with corresponding MC calculations.

摘要

本研究展示了新型selectSeed 125I前列腺籽源与seedSelectron后装设备配套使用时的实验剂量学结果,这符合美国医学物理师协会的建议,即所有新型低能组织间近距离治疗籽源都应进行一次蒙特卡罗(MC)模拟和至少一次实验剂量学特性分析。使用120根经6 MV光子束校准的圆柱形LiF TLD - 100型热释光剂量计进行热释光剂量学测量。将它们置于固体水模体中进行辐照,以实验测定籽源剂量率常数、径向剂量函数和各向异性函数。进行MC模拟以确定热释光剂量计的相对能量响应(发现其与位置无关,且等于1.40±0.03),并计算液态水中剂量与固体水中剂量的比值(发现Dliquidwater/Dsolidwater = 1.013*r + 0.030能很好地描述该比值,且仅对极角有轻微依赖)。发现selectSeed籽源在液态水中的剂量率常数等于0.938±0.065 cGy h(-1) U(-1),在实验不确定度范围内与相应的MC结果lambdaselect Seed = 0.954±0.005 cGy h(-1) U(-1)一致。实验得到的径向剂量和各向异性函数结果也与相应的MC计算结果高度吻合。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验