Robert Normand, Komljenovic Philip T, Rowlands J A
Canadian Heart Research Centre, Toronto, Ontario, Canada.
Med Phys. 2002 May;29(5):736-47. doi: 10.1118/1.1470205.
A method to significantly reduce the exposure area product in fluoroscopy using a pre-patient region-of-interest (ROI) attenuator is presented. The attenuator has a thin central region and a gradually increasing thickness away from the center. It is shown that the unwanted brightening artifact caused by the attenuator can be eliminated by attenuating the low spatial frequencies in the detected image using digital image processing techniques. An investigation of the best image processing method to correct for the presence of the attenuator is undertaken. The correction procedure selected is suitable for use with real-time image processors and the ROI attenuator can be permitted to move during image acquisition. Images of an anthropomorphic chest phantom acquired in the presence of the ROI attenuator using an x-ray image intensifier/video chain are corrected to illustrate the clinical feasibility of our approach.