[Effect of a double intervertebral cage on the mechanical behavior of the lumbar spine].

作者信息

Rohlmann A, Zander Th, Fehrmann M, Klöckner C, Bergmann G

机构信息

Universitätsklinikum Benjamin Franklin, Freie Universität Berlin.

出版信息

Biomed Tech (Berl). 2002 May;47(5):124-9. doi: 10.1515/bmte.2002.47.5.124.

Abstract

A three-dimensional nonlinear finite element model of the lumbar spine was developed. Paired threaded cages and a monosegmental internal spinal fixation device were integrated into the computer model. The model was loaded with such forces as apply during standing, as well as with pure moments in the three main anatomical planes, plus an additional preload. The latter was generated by shortening the distance between the pedicle screws on the longitudinal rod of the fixator. With the exception of torsional loading, an implant appreciably reduces the mobility in the segment concerned. At the loads studied, cages had only a minor impact on the movements and stresses in the adjacent regions, but a strong influence on the stresses in the endplates in contact with them. A preload increases these stresses dramatically. Contact conditions between vertebral body and cages also have a marked effect on the stress distribution in the corresponding vertebral endplate, especially in the case of extension loading. Owing to the preload, maximum stresses were higher for the rigid bond than when contact elements were used.