Gáspár Attila, Andrási Melinda, Kardos Szilvia
Department of Inorganic and Analytical Chemistry, University of Debrecen, H-4010 Debrecen, POB. 21, Hungary.
J Chromatogr B Analyt Technol Biomed Life Sci. 2002 Aug 5;775(2):239-46. doi: 10.1016/s1570-0232(02)00311-2.
The applicability of capillary zone electrophoresis (CZE) for the determination of cephalosporin antibiotics has been studied. In the case of the separation conditions optimised for fourteen cephalosporins, the precision of migration times was smaller than 1.3% RSD, and the values of the limit of detection ranged between 0.42 and 1.62 microg/ml. The proposed CZE method was applied to study the stability of cephalosporins in water at different temperatures (+25, +4 and -18 degrees C). It was established that the degradation of most cephalosporins was not higher than 20% at room temperature within 4 h of dissolution of these antibiotics.