van Gelder Berry M, Bracke Frank A, Meijer Albert
Department of Cardiology, Catharina Hospital, Eindhoven, The Netherlands.
Pacing Clin Electrophysiol. 2002 Jun;25(6):998-9. doi: 10.1046/j.1460-9592.2002.00998.x.
After lead positioning during implantation of a biventricular pacing system intermittent ventricular simulation was observed. There was no pacemaker connected to any of the leads, but ventricular capture could be provoked by an intermittent contact between the lead stylet, which was still partially inserted, and a self-retaining retractor. The same phenomenon was observed from the atrial lead.
在双心室起搏系统植入过程中进行导线定位后,观察到间歇性心室模拟。没有起搏器连接到任何一根导线,但仍部分插入的导线探条与自持牵开器之间的间歇性接触可诱发心室夺获。从心房导线也观察到了同样的现象。