Kreisler Matthias, Al-Haj Haitham, D'Hoedt Bernd
Johannes Gutenberg University, Mainz, Germany.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2002 Jun;93(6):730-5. doi: 10.1067/moe.2002.124766.
The aim of this study was to explore, in vitro, whether the irradiation of human root surfaces with a diode laser might induce nonphysiologic intrapulpal temperature elevations and, therefore, jeopardize pulp vitality.
The pulps were removed from human maxillary and mandibular incisors extracted for periodontal reasons. The root canals were enlarged to an apical size #60 file. The teeth were radiographed with standard dental films and a millimeter grid to determine root thickness. The thickness of dentin between the root surface and the pulp in the irradiation areas was 1, 2, and 3 mm. To determine intrapulpal temperature changes during laser irradiation, 0.5-mm K-type thermocouples were inserted. An 809 nm GaAlAs laser with a 400-micron optical fiber was used. The power output varied between 0.5 and 2.5 W in the continuous-wave mode (0 Hz). Irradiation was continued for up to 120 seconds.
Temperature elevations between 0.5 and 32.0 degrees C were registered in an energy- and time-dependent manner. Dentin thickness had a significant effect on intrapulpal temperature changes (Mann Whitney U test, P <.05), with a thinner dentin layer resulting in higher temperature elevations.
Diode-laser irradiation may jeopardize pulp vitality. It must be recommended to limit power output to 0.5 W and the time of irradiation to 10 seconds when lasing the root surfaces of lower incisors and first maxillary premolars. With other teeth, a power output of 1.0 W and an exposure time of 10 seconds must not be exceeded to ensure a safe clinical application.
本研究的目的是在体外探究,用二极管激光照射人牙根表面是否会引起非生理性的牙髓腔内温度升高,从而危及牙髓活力。
从因牙周原因拔除的人上颌和下颌切牙中取出牙髓。将根管扩大至根尖尺寸#60锉。用标准牙科胶片和毫米网格对牙齿进行X线摄影,以确定牙根厚度。照射区域牙根表面与牙髓之间的牙本质厚度分别为1毫米、2毫米和3毫米。为了确定激光照射期间牙髓腔内的温度变化,插入了0.5毫米的K型热电偶。使用带有400微米光纤的809纳米砷化镓铝激光。在连续波模式(0赫兹)下,功率输出在0.5至2.5瓦之间变化。照射持续长达120秒。
记录到温度升高0.5至32.0摄氏度,呈能量和时间依赖性。牙本质厚度对牙髓腔内温度变化有显著影响(曼-惠特尼U检验,P<.05),牙本质层越薄,温度升高越高。
二极管激光照射可能危及牙髓活力。当对下颌切牙和上颌第一前磨牙的牙根表面进行激光照射时,建议将功率输出限制在0.5瓦,照射时间限制在10秒。对于其他牙齿,为确保安全的临床应用,功率输出不得超过1.0瓦,照射时间不得超过10秒。