Flohr T, Bruder H, Stierstorfer K, Simon J, Schaller S, Ohnesorge B
Siemens Medical Solutions, Computed Tomography, Forchheim, Germany.
Rofo. 2002 Aug;174(8):1022-7. doi: 10.1055/s-2002-32930.
Despite all promising advances, some challenges remain for ECG-gated multislice CT examinations of the heart and the coronary arteries with current 4-slice detectors: adequate visualization of stents and severely calcified coronary arteries, examination of patients with higher heart rates and patients, who cannot adequately hold their breath for at least 30 sec. The new generation of multislice CT systems offering simultaneous acquisition of up to 16 sub-millimeter slices and gantry rotation times shorter than 0.5 sec has the potential to overcome these limitations. We describe the technical principles of cardiac scanning with a state-of-the-art 16-slice CT equipment (SOMATOM Sensation 16, Siemens AG, Forchheim, Germany). We discuss an extension of the Adaptive Cardio Volume (ACV) reconstruction approach for ECG-gated multislice spiral CT. We show the impact of reduced gantry rotation time (0.42 sec) on temporal resolution, and we demonstrate the influence of slice width on the visualization of stents and plaques. Deviating from general purpose applications a cone-correction is not required for cardiac scanning with 16-slice CT systems. In addition to the absolute improvement, the temporal resolution shows a different dependence on the patient's heart rate for 0.42 sec rotation time, reaching its optimum (105 msec) at 81 BPM. This has the potential to expand the range of heart rates accessible to routine clinical examinations. Owing to sub-millimeter slice width and optimized in-plane resolution characteristics, visualization of stents and severe calcifications in coronary arteries is significantly improved. Clinical experience will be needed to fully evaluate the potential of 16-slice technology for cardiac imaging.
尽管取得了所有这些令人鼓舞的进展,但对于使用当前的4层探测器对心脏和冠状动脉进行心电图门控多层CT检查而言,仍存在一些挑战:支架和严重钙化的冠状动脉的可视化效果欠佳、心率较高的患者以及无法屏气至少30秒的患者的检查。新一代多层CT系统能够同时采集多达16层亚毫米级切片,且机架旋转时间短于0.5秒,有可能克服这些限制。我们描述了使用最先进的16层CT设备(西门子公司的SOMATOM Sensation 16,德国福希海姆)进行心脏扫描的技术原理。我们讨论了用于心电图门控多层螺旋CT的自适应心脏容积(ACV)重建方法的扩展。我们展示了缩短机架旋转时间(0.42秒)对时间分辨率的影响,并证明了切片宽度对支架和斑块可视化的影响。与通用应用不同,使用16层CT系统进行心脏扫描不需要进行锥形校正。除了绝对的改善之外,对于0.42秒的旋转时间,时间分辨率对患者心率的依赖性也有所不同,在心率为81次/分钟时达到最佳值(105毫秒)。这有可能扩大常规临床检查可触及的心率范围。由于亚毫米级切片宽度和优化的平面分辨率特性,冠状动脉中支架和严重钙化的可视化效果得到显著改善。需要临床经验来全面评估16层技术在心脏成像方面的潜力。