Rajendran V, Lrudayaraj J
Department of Agricultural and Biological Engineering, The Pennsylvania State University, University Park 16802, USA.
J Dairy Sci. 2002 Jun;85(6):1357-61. doi: 10.3168/jds.s0022-0302(02)74201-x.
A microdialysis-coupled flow injection amperometric Sensor (microFIAS) was used to determine glucose, galactose, and lactose in milk. The sensor is based on enzyme-catalyzed reaction in combination with the three well-established analytical techniques, namely; microdialysis sampling, flow injection analysis (FIA), and amperometric detection. With the multianalyte sensor it was possible to detect glucose and galactose by sequential injection of their corresponding oxidase enzymes: glucose oxidase and galactose oxidase, while lactose was determined by injection of a mixture of beta-galactosidase and glucose oxidase enzymes. The sensor showed a linear response between 0.05 and 10 mM for glucose, between 0.1 and 20 mM for galactose and between 0.2 and 20 mM for lactose, respectively. The relative standard deviation values of the sensor measurements for glucose, galactose, and lactose were 3-4% (n = 3). The sensor measurements for lactose content in milk were compared with a standard method with an infrared spectrophotometer.
一种微透析耦合流动注射安培传感器(microFIAS)用于测定牛奶中的葡萄糖、半乳糖和乳糖。该传感器基于酶催化反应,并结合了三种成熟的分析技术,即微透析采样、流动注射分析(FIA)和安培检测。使用这种多分析物传感器,可以通过依次注入其相应的氧化酶(葡萄糖氧化酶和半乳糖氧化酶)来检测葡萄糖和半乳糖,而乳糖则通过注入β-半乳糖苷酶和葡萄糖氧化酶的混合物来测定。该传感器对葡萄糖的线性响应范围为0.05至10 mM,对半乳糖为0.1至20 mM,对乳糖为0.2至20 mM。葡萄糖、半乳糖和乳糖传感器测量的相对标准偏差值为3-4%(n = 3)。将牛奶中乳糖含量的传感器测量结果与红外分光光度计的标准方法进行了比较。