接触式二极管激光经巩膜睫状体光凝术中总能量传递对成功率的影响:一项回顾性病例分析与荟萃分析

Influence of total energy delivery on success rate after contact diode laser transscleral cyclophotocoagulation: a retrospective case review and meta-analysis.

作者信息

Hauber Frederick A, Scherer Warren J

机构信息

Pasco Eye Institute, New Port Richey, Florida 34652, USA.

出版信息

J Glaucoma. 2002 Aug;11(4):329-33. doi: 10.1097/00061198-200208000-00009.

Abstract

PURPOSE

To evaluate the effect of a single treatment of contact diode laser transscleral cyclophotocoagulation (TSCP) on intraocular pressure (IOP) in patients with refractory glaucoma and perform a historical meta-analysis of the relationship between successful IOP control and various laser parameters.

METHODS

Clinical data for 47 eyes of 43 patients who underwent a single session of contact diode laser TSCP (power of 2000 mW, exposure time of 2 seconds and 25.6 average applications) for refractory glaucoma at Pasco Eye Institute between 1996 and 2001 were reviewed. Baseline IOP was compared with the postoperative IOP at 1 month, 3 to 6 months, and 1 year. Snellen visual acuity and the number of glaucoma medications being used 3 to 6 months postoperatively were compared with baseline. Complications were also evaluated. The relationship between successful IOP control and various laser parameters (energy applied per eye, energy per laser application, power, exposure time and number of laser applications) was examined using the results of this study and previous investigations.

RESULTS

The mean +/- standard deviation (SD) baseline IOP was 29.4 +/- 10.9 mm Hg. Diode laser TSCP decreased average IOP to 15.7 +/- 12.0 mm Hg at 1 month, 16.3 +/- 4.2 mm Hg at 3 to 6 months, and 16.2 +/- 4.8 mm Hg at 1 year, postoperatively. This represents an IOP reduction of approximately 45%. The percentage of patients maintaining a postoperative IOP of <21 mm Hg was 95.7% at 1 month, 92.1% at 3 to 6 months, and 94.4% at 1 year after TSCP. At the 3 to 6 month postoperative visit, the average +/- SD change in Snellen visual acuity (-0.5 +/- 1.1 lines) and the number of glaucoma medications used did not differ significantly from baseline. Combined analysis of the results of this study and previous studies reveals a linear, direct correlation (r = 0.91) between the percentage of patients achieving a successful outcome (defined as a final IOP < 21 or 22 mm Hg) and the total energy delivered to the ciliary body during the treatment session.

CONCLUSIONS

Diode laser TSCP is a safe and highly effective method for lowering IOP in patients with uncontrolled glaucoma. Higher total energy levels in a given treatment session appear to be associated with increased success as defined by the percentage of patients achieving an IOP < 21 or 22 mm Hg, without an increased risk of additional complications or vision loss.

摘要

目的

评估单次接触式二极管激光经巩膜睫状体光凝术(TSCP)对难治性青光眼患者眼压(IOP)的影响,并对眼压成功控制与各种激光参数之间的关系进行历史荟萃分析。

方法

回顾了1996年至2001年间在帕斯科眼科研究所对43例难治性青光眼患者的47只眼睛进行单次接触式二极管激光TSCP治疗(功率2000 mW,曝光时间2秒,平均应用25.6次)的临床数据。将基线眼压与术后1个月、3至6个月和1年时的眼压进行比较。将术后3至6个月时的Snellen视力和使用的青光眼药物数量与基线进行比较。还对并发症进行了评估。利用本研究结果和既往研究结果,研究了眼压成功控制与各种激光参数(每只眼睛施加的能量、每次激光应用的能量、功率、曝光时间和激光应用次数)之间的关系。

结果

平均±标准差(SD)基线眼压为29.4±10.9 mmHg。二极管激光TSCP术后1个月时平均眼压降至15.7±12.0 mmHg,3至6个月时降至16.3±4.2 mmHg,1年时降至16.2±4.8 mmHg。这代表眼压降低了约45%。TSCP术后1个月时,术后眼压维持在<21 mmHg的患者百分比为95.7%,3至6个月时为92.1%,1年时为94.4%。在术后3至6个月的随访中,Snellen视力的平均±SD变化(-0.5±1.1行)和使用的青光眼药物数量与基线相比无显著差异。对本研究结果和既往研究结果的综合分析显示,达到成功结局(定义为最终眼压<21或22 mmHg)的患者百分比与治疗期间传递至睫状体的总能量之间存在线性直接相关性(r = 0.91)。

结论

二极管激光TSCP是降低未控制青光眼患者眼压的一种安全且高效的方法。在给定治疗疗程中,较高的总能量水平似乎与更高的成功率相关,成功的定义为眼压<21或22 mmHg的患者百分比增加,且不会增加额外并发症或视力丧失的风险。