Optimum iodine concentration of contrast material through microcatheters: hydrodynamic analysis of experimental results.

作者信息

Yamauchi Teiyu, Furui Shigeru, Harasawa Arimi, Ishimura Makoto, Imai Tatsuhiko, Hayashi Toshimitsu

机构信息

Department of Radiology, Teikyo University School of Medicine, Tokyo, Japan.

出版信息

Phys Med Biol. 2002 Jul 21;47(14):2511-23. doi: 10.1088/0031-9155/47/14/310.

Abstract

It is important to increase the iodine delivery rate (I), that is the iodine concentration of the contrast material (C) x the flow rate of the contrast material (Q), through a catheter to obtain high quality arteriograms. The iodine delivery rate varies depending on the iodine concentration of the contrast material. The purpose of this study is to estimate the optimum iodine concentration (Copt) of contrast material (i.e. the iodine concentration at which I becomes maximum) through a microcatheter of a given length (L), inner diameter (D) and injection pressure (P). Iohexol, ioversol and iopamidol of 11-12 iodine concentrations (140-350, 160-350 or 160-370 mg cm(-3)) at 37 degrees C are used. I and Reynolds number (Re) of the flow of each contrast material through four microcatheters (0.49-0.68 mm in inner diameter, 1000-1500 mm in length) at injection pressures of 1.38, 2.76, 4.14 and 5.52 x 10(6) Pa (200, 400, 600 and 800 pounds per square inch) are obtained experimentally. The relationships between I and C and between I and Re are examined for each catheter and injection pressure. Copt is 160-280 mg cm(-3) for iohexol, 180-280 mg cm(-3) for ioversol and 200-300 mg cm(-3) for iopamidol. I becomes maximum when Re approximates the critical Reynolds number (Re approximately 2300). Utilizing this principle, we can estimate Copt and its flow rate through a microcatheter of a given L, D and P.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验