Moreno-Montañés J, Alvarez-Vidal A, Sainz Gómez C, Rodríguez Conde R
Departamento de Oftalmología, Clínica Universitaria, Universidad de Navarra, Pamplona, España.
Arch Soc Esp Oftalmol. 2002 Aug;77(8):435-41.
To investigate the retinal nerve fiber layer (RNFL) thickness in glaucomatous eyes using Optical Coherence Tomography (OCT). To compare the RNFL thickness with visual field damage (Humphrey field analyzer).
The mean RNFL thickness in glaucomatous eyes (n = 80) was compared with age-matched normal eyes (n = 40). Three circular scans were obtained for each eye using OCT (3.4 mm diameter). In each eye, average RNFL, 4 quadrants and 12 meridians were calculated and compared. The superior-inferior asymmetry of RNFL was studied. The average RNFL thickness was compared with mean deviation (MD) and mean standard deviation (MSD) of SITA 24-2 visual field program.
Mean RNFL was significantly thinner in glaucomatous eyes than in normal eyes (p<0.001). The RNFL thickness was decrease in the 4 quadrants and 11 of 12 meridians studied in glaucomatous eyes (p<0.05). The superior-inferior asymmetry showed a significant difference in RNFL thickness at 30 degrees central meridian (p<0.05). Mean RNFL thickness was significantly associated with DM and DSM of visual field (p<0.001) in glaucomatous eyes.
OCT revealed significant quantitative differences in RNFL thickness between glaucomatous and normal eyes. OCT showed a considerable measurements overlap between glaucomatous and normal eyes, which can limit the sensitivity and specificity of this instrument (Arch Soc Esp Oftalmol 2002; 77: 435-442).
使用光学相干断层扫描(OCT)研究青光眼患者视网膜神经纤维层(RNFL)厚度。将RNFL厚度与视野损害情况(Humphrey视野分析仪)进行比较。
将青光眼患者(n = 80)的平均RNFL厚度与年龄匹配的正常受试者(n = 40)进行比较。使用OCT(直径3.4 mm)对每只眼睛进行三次环形扫描。计算并比较每只眼睛的平均RNFL、四个象限及12条子午线的情况。研究RNFL的上下不对称性。将平均RNFL厚度与SITA 24-2视野检查程序的平均偏差(MD)和平均标准差(MSD)进行比较。
青光眼患者的平均RNFL厚度显著低于正常受试者(p<0.001)。青光眼患者研究的四个象限及12条子午线中的11条的RNFL厚度降低(p<0.05)。上下不对称性显示在中央子午线30度处RNFL厚度存在显著差异(p<0.05)。青光眼患者的平均RNFL厚度与视野的MD和DSM显著相关(p<0.001)。
OCT显示青光眼患者与正常受试者之间RNFL厚度存在显著的定量差异。OCT显示青光眼患者与正常受试者之间存在相当程度的测量重叠,这可能会限制该仪器的敏感性和特异性(《西班牙眼科协会杂志》2002年;77:435 - 442)。