Rivard Mark J
Department of Radiation Oncology, Tufts University School of Medicine, New England Medical Center #246, 750 Washington Street, Boston, MA 02111, USA.
Appl Radiat Isot. 2002 Sep;57(3):381-9. doi: 10.1016/s0969-8043(02)00110-0.
The Model 3500 I-Plant 125I brachytherapy source, produced by Implant Sciences Corporation, is created through a novel technique which utilizes ion implantation of 124Xe into a ceramic matrix followed by neutron activation of the substrate. Clinical dosimetry parameters were calculated using MCNP. For data processing, a 0.376 cm active length was used for the geometry function. The 2-D anisotropy function, F(r, 0), exhibited less anisotropy than the Amersham Health Model 6711 source, and decreased as radial distance, r, increased. The anisotropy constant, phi an, was 0.933; dose rate constant, lambda, was 1.017 +/- 0.005 cGy/h/U; and radial dose function, g(r), for r = (0.1, 0.2, 0.5, 2, 5, 10 cm) was 0.990, 1.014, 1.030, 0.872, 0.448, and 0.114, respectively. In general, there was good agreement with experimental results recently obtained by others using thermoluminescent dosimeters.
由植入科学公司生产的3500 I型植物125I近距离放射治疗源,是通过一种新技术制造的,该技术利用将124Xe离子注入陶瓷基质,然后对基质进行中子活化。使用MCNP计算临床剂量学参数。在数据处理中,几何函数使用0.376厘米的活性长度。二维各向异性函数F(r, 0)表现出比阿美仙健康公司6711型源更小的各向异性,并且随着径向距离r的增加而减小。各向异性常数φan为0.933;剂量率常数λ为1.017±0.005 cGy/h/U;对于r = (0.1, 0.2, 0.5, 2, 5, 10厘米),径向剂量函数g(r)分别为0.990、1.014、1.030、0.872、0.448和0.114。总体而言,与其他研究人员最近使用热释光剂量计获得的实验结果吻合良好。