Preparation of mixtures of humidified air with CO, SO2, NO2, NH3, NH2CH3, Cl2, and HCl to measure respirator cartridge and canister service life.

作者信息

Nelson G O, Swisher L E, Taylor R D, Bigler B E

出版信息

Am Ind Hyg Assoc J. 1975 Jan;36(1):49-56. doi: 10.1080/0002889758507207.

Abstract

Four systems have been fabricated to measure the effective service life of respirator cartridges and canisters. The four systems employ carbon monoxide, sulfur dioxide/nitrogen dioxide, ammonia/methylamine, and chlorine/hydrogen chloride as contaminants. Each system produces test concentrations in the low percentage range (typically 0.05 to 3.0%) at flow rates of 32 and 64 liters/min and at relative humidities between 20 and 95%. Suitable anlyzers have been provided to measure the downstream concentration from 1 to 500 ppm as a function of time. Holders have been designed to accommodate the wide size variation of commercially available cartridges and canisters.