平板数字X线摄影与存储磷光体计算机X线摄影的比较:体模研究中剂量与图像质量的评估

Flat panel digital radiography compared with storage phosphor computed radiography: assessment of dose versus image quality in phantom studies.

作者信息

Fischbach Frank, Ricke Jens, Freund Torsten, Werk Michael, Spors Birgit, Baumann Clemens, Pech Marcie J, Felix Roland

机构信息

Department of Radiology, Charité Virchow-Klinikum, Medical Faculty of the Humboldt University Berling, Germany.

出版信息

Invest Radiol. 2002 Nov;37(11):609-14. doi: 10.1097/00004424-200211000-00004.

Abstract

RATIONALE AND OBJECTIVE

To assess and quantify the dose reduction by use of a CsI-flat panel digital radiography (DR)-system compared with digital computed radiography (CR).

MATERIALS AND METHODS

A TCDD-test using the CDRAD-phantom was performed at mAs-values of 5, 4, 2.5, 2, 1, and 0.5 mAs for both digital systems. Entrance surface doses were recorded for all images. Images were presented to four independent observers. For quantitative comparison the image quality figure (IQF) was calculated. Statistical analysis was performed using the Pearson correlation and the Wilcoxon test. A ROC analysis was performed using the TRG-phantom. Settings of 4, 2.5, 2 mAs for both systems were used. In addition, 1 and 0.5 mAs were used for the DR system only. Statistical significance was evaluated using Student test.

RESULTS

The DR system provided equivalent results compared with CR with respect to high frequency information and superior results with respect to low contrast details. Compared with computed radiography, the flat panel detector demonstrated significantly lower IQFs, ensuring a better image quality with respect to contrast and detail detectability. IQFs for DR and CR were equal at a surface dose reduction of 87% for DR. ROC analysis revealed significantly higher values under the curve for DR up to a surface dose reduction of 70%.

CONCLUSIONS

Image quality of DR proved to be far superior to CR in particular for low contrast details. The image quality of CR is similar to that of DR only at high dose levels.

摘要

原理与目的

评估并量化使用碘化铯平板数字射线摄影(DR)系统相较于数字计算机射线摄影(CR)的剂量降低情况。

材料与方法

使用CDRAD体模进行TCDD测试,两个数字系统的管电流-时间乘积(mAs)值分别为5、4、2.5、2、1和0.5 mAs。记录所有图像的体表剂量。将图像呈现给四位独立观察者。为进行定量比较,计算图像质量因子(IQF)。使用Pearson相关性分析和Wilcoxon检验进行统计分析。使用TRG体模进行ROC分析。两个系统均采用4、2.5、2 mAs的设置。此外,仅对DR系统使用1和0.5 mAs。使用Student检验评估统计学显著性。

结果

在高频信息方面,DR系统与CR相比提供了等效结果,在低对比度细节方面结果更优。与计算机射线摄影相比,平板探测器的IQF显著更低,在对比度和细节可探测性方面确保了更好的图像质量。当DR的体表剂量降低87%时,DR和CR的IQF相等。ROC分析显示,在体表剂量降低70%之前,DR的曲线下值显著更高。

结论

DR的图像质量被证明远优于CR,特别是在低对比度细节方面。仅在高剂量水平时,CR的图像质量与DR相似。