Wang Hanwen, Xu Hui, Guan Yafeng
Department of Analytical Chemistry and Micro-Instrumentation, Dalian Institute of Chemical Physics, Chinese Academy of Sciences.
J Chromatogr A. 2002 Oct 4;972(2):253-8. doi: 10.1016/s0021-9673(02)01090-7.
A simple method, based on the technique of capillary column switching-back flushing, has been developed for the detailed analysis of aromatic compounds in gasoline. The sample was first separated on a 30-m long OV-2330 polar precolumn and then backflushed onto a nonpolar analytical column. The early eluting components from the precolumn and the components of interest (aromatic compounds plus heavier compounds) eluting from the analytical column are all directed to the same flame ionization detection system through a T piece, which permits the quantitative analysis of aromatic hydrocarbons in gasoline by a normalization method using correcting factors. The switching time window of the method is +/-5 s, resulting in easier operation and higher reliability. The reproducibility of the quantitative analysis was < or = 3% RSD for real gasoline samples.
基于毛细管柱切换-反吹技术,已开发出一种用于详细分析汽油中芳香族化合物的简单方法。样品首先在一根30米长的OV-2330极性预柱上进行分离,然后反吹到一根非极性分析柱上。预柱上早期洗脱的组分以及分析柱上洗脱的目标组分(芳香族化合物和较重的化合物)都通过一个三通管导向同一个火焰离子化检测系统,这使得可以通过使用校正因子的归一化方法对汽油中的芳烃进行定量分析。该方法的切换时间窗口为±5秒,操作更简便,可靠性更高。对于实际汽油样品,定量分析的重现性RSD≤3%。