Stout Peter R, Gehlhausen Jay M, Horn Carl K, Klette Kevin L
Navy Drug Screening Laboratory, Naval Air Station, Jacksonville, Florida 32212, USA.
J Anal Toxicol. 2002 Oct;26(7):401-5. doi: 10.1093/jat/26.7.401.
A novel extraction and derivatization procedure for the cocaine metabolite benzoylecgonine (BZE) was developed and evaluated for use in a high-volume forensic urine analysis laboratory. Extractions utilized a Speedisk 48 positive pressure extraction manifold and polymer-based cation-exchange extraction columns. Samples were derivatized by the addition of pentafluoropropionic anhydride and pentafluoropropanol. All analyses were performed in selected ion monitoring mode; ions included m/z 421, 300, 272, 429, and 303 with m/z 421 to 429 ratio used for quantitation. The average extraction efficiency was 80%. Seventy-five common over-the-counter products, including prescription drugs, drug metabolites, and other drugs of abuse, demonstrated no significant interference with respect to chromatography or quantitation. The limit of detection and limit of quantitation were calculated at 12.5 ng/mL, and the assay was linear from 12.5 to 20,000 ng/mL with an r2 of 0.99932. A series of 20 precision samples (100 ng/mL) produced an average response of 97.8 ng/mL and a percent coefficient of variation of 4.1%. A set of 79 archived human urine samples that had previously been found to contain BZE were analyzed by 3 separate laboratories. The results did not differ significantly from prior quantitation or between laboratories. The Speedisk has proven viable for a high-volume production facility reducing overall cost of analysis by decreasing analysis time and minimizing waste production while meeting strict forensic requirements.
开发并评估了一种用于可卡因代谢物苯甲酰芽子碱(BZE)的新型提取和衍生化方法,以用于大容量法医尿液分析实验室。提取过程使用了Speedisk 48正压提取歧管和基于聚合物的阳离子交换提取柱。通过添加五氟丙酸酐和五氟丙醇对样品进行衍生化。所有分析均在选择离子监测模式下进行;离子包括m/z 421、300、272、429和303,以m/z 421与429的比率用于定量。平均提取效率为80%。75种常见的非处方药产品,包括处方药、药物代谢物和其他滥用药物,在色谱或定量方面均未显示出明显干扰。检测限和定量限计算为12.5 ng/mL,该测定在12.5至20,000 ng/mL范围内呈线性,r2为0.99932。一系列20个精密度样品(100 ng/mL)的平均响应值为97.8 ng/mL,变异系数百分比为4.1%。由3个独立实验室对一组79份先前已被发现含有BZE的存档人类尿液样本进行了分析。结果与先前的定量结果或不同实验室之间的结果无显著差异。事实证明,Speedisk对于大容量生产设施是可行的,它通过减少分析时间和将废物产生降至最低,同时满足严格的法医要求,降低了总体分析成本。