Kadis Rouvim
D.I. Mendeleyev Institute for Metrology, 19 Moskovsky pr., 198005 St Petersburg, Russia.
Anal Bioanal Chem. 2002 Nov;374(5):817-23. doi: 10.1007/s00216-002-1457-x. Epub 2002 Sep 18.
Establishing a traceability route with all measurement and uncertainty relationships determined is an important aspect of traceability, and seems to be particularly striking in pH measurement. In this paper the issue of evaluation of secondary pH standards measured with reference to a primary standard in a differential cell with free diffusion type liquid junctions is considered. Relatively high uncertainty, U=0.015, has been assigned to such standards in the recent IUPAC Recommendations on pH (2001), because of a specific residual liquid-junction potential treated statistically as a contribution to the combined uncertainty. Close inspection of the problem leads to the conclusion that a correction for the residual liquid-junction potential should be applied to the measured value of a secondary pH standard. This can be considered as a correction for a known systematic effect on the traceability route. With available experimental data it is demonstrated that such a correction can reasonably be made for well-studied standard buffer systems. In this way the uncertainty associated with secondary pH standards is kept to a low level, and, what is more, the problem of two pH scales, a multi-standard scale and a single-standard scale, gains a proper solution. The need for different treatment of residual liquid-junction potentials at different levels in the measurement hierarchy is noted. Much attention is also given to rational categorization of pH standards in the hierarchy.
建立一条确定了所有测量和不确定度关系的可追溯路径是可追溯性的一个重要方面,这在pH测量中似乎尤为突出。本文考虑了在具有自由扩散型液接界的微分电池中,相对于一级标准测量二级pH标准的评估问题。由于将特定的残余液接界电势统计地视为合成不确定度的一个贡献,在最近的IUPAC关于pH的建议(2001年)中,此类标准被赋予了相对较高的不确定度U = 0.015。对该问题的仔细研究得出结论,应对二级pH标准的测量值应用残余液接界电势的校正。这可被视为对可追溯路径上已知系统效应的校正。利用现有的实验数据表明,对于经过充分研究的标准缓冲体系,可以合理地进行这种校正。通过这种方式,与二级pH标准相关的不确定度被保持在较低水平,而且,更重要的是,多标准刻度和单标准刻度这两种pH刻度的问题得到了妥善解决。文中指出了在测量层次结构的不同级别对残余液接界电势进行不同处理的必要性。同时也非常关注pH标准在层次结构中的合理分类。