Zhang Yongchao, Kim Hyug-Han, Mano Nicolas, Dequaire Murielle, Heller Adam
Department of Chemical Engineering and Texas Materials Institute, The University of Texas at Austin, Austin, Texas 78712-1062, USA.
Anal Bioanal Chem. 2002 Nov;374(6):1050-5. doi: 10.1007/s00216-002-1604-4. Epub 2002 Oct 15.
A 38-base DNA sequence has been detected at 20 pmol L(-1) concentration in 15-35- microL droplets by means of an electrochemical enzyme-amplified sandwich-type assay on a mass-manufacturable screen-printed carbon electrode. Formation of the sandwich brought the horseradish peroxidase-label of the detection sequence into electrical contact with a pre-electrodeposited redox polymer, making the sandwich an electrocatalyst for the reduction of hydrogen peroxide to water at +0.2 V (Ag/AgCl). Sensitivity twenty times better than that of a related system resulted from: 1. fivefold reduction of the noise by substituting the formerly used poly( N-vinyl imidazole)-co-acrylamide comprising redox co-polymer with poly(4-vinyl pyridine)-co-acrylamide comprising redox polymer, enabling use of the electrodes at a more oxidizing potential at which noise (the rate of non-enzyme catalyzed electroreduction currents of dissolved oxygen and hydrogen peroxide) was lower; 2. doubling of the catalytic electroreduction current upon electrodeposition of a second layer of the redox polymer on the capture sequence-containing film; and 3. doubling of the current by increasing the coverage by the capture sequence.
通过在可大规模生产的丝网印刷碳电极上进行电化学酶放大夹心型测定法,在浓度为20 pmol L(-1)的15 - 35微升液滴中检测到了一段38个碱基的DNA序列。夹心结构的形成使检测序列的辣根过氧化物酶标记物与预电沉积的氧化还原聚合物实现电接触,使得该夹心结构成为在+0.2 V(Ag/AgCl)下将过氧化氢还原为水的电催化剂。灵敏度比相关系统提高了二十倍,原因如下:1. 用包含氧化还原聚合物的聚(4 - 乙烯基吡啶)- 共 - 丙烯酰胺替代先前使用的包含氧化还原共聚物的聚(N - 乙烯基咪唑)- 共 - 丙烯酰胺,使噪声降低了五倍,从而能够在氧化电位更高的情况下使用电极,此时噪声(溶解氧和过氧化氢的非酶催化电还原电流速率)更低;2. 在含有捕获序列的膜上电沉积第二层氧化还原聚合物后,催化电还原电流增加了一倍;3. 通过增加捕获序列的覆盖率使电流增加了一倍。