Saito Yoshihiro, Nojiri Masatoshi, Imaizumi Motohiro, Nakao Yuji, Morishima Yoriko, Kanehara Hideyuki, Matsuura Hisashi, Kotera Kenzo, Wada Hiroo, Jinno Kiyokatsu
School of Materials Science, Toyohashi University of Technology, Toyohashi 441-8580, Japan.
J Chromatogr A. 2002 Oct 25;975(1):105-12. doi: 10.1016/s0021-9673(02)01333-x.
Miniaturized sample preparation technique for complex sample matrices has been developed with a polymer-coated fibrous extraction medium. Several hundreds of fine fibrous materials were packed longitudinally into a fused-silica capillary followed by a polymeric coating on it to prepare the extraction capillary. The extraction capillary was installed in a liquid chromatograph as a sample loop of the injection valve. The on-line coupled sample preparation/separation system demonstrated a good validity for the analysis of phthalates in real river and wastewater samples. The lowest limits of quantification for several phthalates were less than 1 ng/ml. The effect of polymeric coating to the filaments on the extraction power was also investigated.