Hammer W J, Hearne M J, Roberts W C
J Thorac Cardiovasc Surg. 1976 Feb;71(2):259-61.
Intermittent aortic regurgitation due to cocking is described for the first time after replacement of the aortic valve with a poppet-disc prosthesis. A combination of disc grooving and strut thrombus produced the cocking with resultant aortic regurgitation.
首次报道了在使用球瓣人工心脏瓣膜置换主动脉瓣后出现的因瓣膜翘起导致的间歇性主动脉瓣反流。瓣叶开槽和支柱血栓共同作用导致瓣膜翘起,进而引起主动脉瓣反流。