Wunderlich W, Fischer F, Arntz H R, Horras N, Schwimmbeck P, Morguet A J, Schultheiss H P
Abteilung Innere Medizin II-Kardiologie und Pulmologie, Universitätsklinikum Benjamin Franklin, Freie Universität Berlin, Deutschland.
Biomed Tech (Berl). 2002;47 Suppl 1 Pt 2:648-51. doi: 10.1515/bmte.2002.47.s1b.648.
In this study a procedure for online measurement of a coronary segment length during primary stenting is presented. The spatial segmental axis is calculated from a biplane angiogram and the corresponding projection data by simple delineation of the proximal and distal segment sites in each image. In a clinical evaluation the length error, projectional foreshortening and time frame during measurement of 222 stents or carrier balloons were calculated. The overall percentage length error was 3.3(+)-2.4%. The device length imaged with a mean projectional foreshortening of 18(+)-20% was measured within 18(+)-4 s. The procedure yields the accurate length of coronary segments within a short computation time taking projectional foreshortening into account.