Suppr超能文献

Direct determination of Cr and Cu in urine samples by electrothermal atomic absorption spectrometry using ruthenium as permanent modifier (R1).

作者信息

Lelis Kátia Linces Alves, Magalhães Cristina Gonçalves, Rocha Clélia Aparecida, of Silva José Bento Borba

机构信息

lnstituto de Patologia Clínica Hermes Pardini, 30140-070 Belo Horizonte, MG, Brazil.

出版信息

Anal Bioanal Chem. 2002 Dec;374(7-8):1301-5. doi: 10.1007/s00216-002-1556-8. Epub 2002 Nov 12.

Abstract

In this study Ru, deposited thermally on an integrated platform pyrolytic graphite tube, is proposed as a permanent modifier for the determination of Cu and Cr in urine samples by electrothermal atomic absorption spectrometry. The samples were diluted 1:1 with nitric acid (1% v/v). Pyrolysis and atomization temperatures for spiked urine samples were 1,100 degrees C and 1,900 degrees C respectively for Cu, and 1,400 degrees C and 2,500 degrees C respectively for Cr. For comparison purposes, the conventional modification with Pd+Mg was also studied. The sensitivity for Ru as permanent modifier was higher for the two analytes. The characteristic masses were 7.3 and 17.7 for Cr and Cu. The detection limits (3sigma) were 0.22 and 0.32 microg/L, for Cr and Cu, respectively. Good agreement was obtained with certified urine samples for the two elements.