Watanabe Tetsu, Tomita Tetsuya, Fujii Masakazu, Kaneko Motoharu, Sakaura Hironobu, Takeuchi Eiji, Sugamoto Kazuomi, Yoshikawa Hideki
Division of Computer Integrated Orthopaedic Surgery, Osaka University Graduate School of Medicine, Suita, Osaka, Japan.
J Arthroplasty. 2002 Dec;17(8):1058-62. doi: 10.1054/arth.2002.35792.
Periprosthetic fracture of the tibial plateau associated with osteolysis resulting from mechanical failure of the rotating patellar component after total knee arthroplasty with the New Jersey Low-Contact-Stress (LCS) knee (DePuy, Warsaw, IN) has not been reported previously. A 67-year-old woman with rheumatoid arthritis of the left knee had a LCS prosthesis implanted without cement, using a rotating patellar component. Seven years later, a fracture of the lateral tibial plateau occurred owing to an osteolytic defect with no traumatic accident. The rotating patellar bearing over-rotated and locked; consequently, wear occurred between the patellar metal tray and the femoral component. Immunohistochemistry revealed CD68-positive macrophages in the osteolytic region and phagocytosis of metal particles. The osteolytic region was filled with autogenous bone, and all components were exchanged and cemented. The patient's condition became satisfactory with relief of pain.
采用新泽西低接触应力(LCS)膝关节(DePuy公司,印第安纳州华沙市)进行全膝关节置换术后,因旋转髌骨组件机械故障导致骨溶解而并发胫骨平台假体周围骨折的情况此前未见报道。一名67岁患有左膝类风湿性关节炎的女性接受了无骨水泥植入的LCS假体置换,使用了旋转髌骨组件。7年后,由于溶骨性缺损且无外伤事故,发生了外侧胫骨平台骨折。旋转髌骨轴承过度旋转并锁定;因此,髌骨金属托盘与股骨组件之间出现磨损。免疫组织化学显示溶骨区域有CD68阳性巨噬细胞以及金属颗粒的吞噬现象。溶骨区域用自体骨填充,所有组件均进行了更换并使用骨水泥固定。患者疼痛缓解,情况令人满意。