Lemaire Roger
Orthopaedic Department, University Hospital, Liège, Belgium.
Acta Orthop Belg. 2010 Apr;76(2):274-81.
Stress fracture of the femoral component is a rare complication following total knee arthroplasty. We report one such complication which occurred 6.5 years after primary implantation of an uncemented meniscal Low Contact Stress (LCS) prosthesis. The fracture affected the medial condyle of the femoral component, and a large osteolytic lesion was present in the underlying femoral condyle. The patient underwent revision arthroplasty with a cemented constrained condylar knee, which is still in place with a satisfactory result 12 years later. This complication appears to be very rare, although a number of cases probably go unreported. From data available in literature, it appears that stress fractures of the femoral component have predominantly affected the medial condyle, following uncemented implantation of fixed-bearing as well as mobile-bearing knees. Different mechanisms may be involved: specific design features of the implant, failure of bone ingrowth in uncemented components, and osteolysis from particulate wear debris.
股骨部件应力性骨折是全膝关节置换术后一种罕见的并发症。我们报告了一例这样的并发症,该并发症发生在初次植入非骨水泥型半月板低接触应力(LCS)假体6.5年后。骨折累及股骨部件的内侧髁,其下方的股骨髁出现一个大的溶骨性病变。患者接受了骨水泥固定型限制性髁膝关节翻修术,12年后该假体仍在位,效果满意。这种并发症似乎非常罕见,尽管可能有一些病例未被报道。从文献中的现有数据来看,在非骨水泥固定型以及活动型膝关节植入后,股骨部件的应力性骨折主要累及内侧髁。可能涉及不同的机制:植入物的特定设计特点、非骨水泥部件骨长入失败以及颗粒磨损碎屑导致的骨溶解。