Kucukosmanoglu Osman, Celiker Alpay, Ozer Sema, Karagoz Tevfik
Department of Pediatric Cardiology, Cukurova University, Adana, Turkey.
Pacing Clin Electrophysiol. 2002 Nov;25(11):1624-7. doi: 10.1046/j.1460-9592.2002.01624.x.
The Autocapture function controls and optimizes the amplitude of the pacing pulse and saves energy. The manufacturer recommends using a special low polarization, low threshold bipolar Pacesetter lead for the Autocapture function. The purpose of this study was to evaluate the compatibility of Autocapture with previously implanted pacing leads. The study included 15 patients (mean age 13.6 +/- 3.4 years) who needed pulse generator replacement and received the VVIR pacemaker Regency SR+ or the DDDR pacemakers Affinity DR or Integrity DR with the Autocapture function. The new pulse generators connected to previously implanted ventricular leads. At the time of implantation the pacing threshold was 1.0 +/- 0.35 V at 0.5 ms, the lead impedance was 580 +/- 80 omega, and the spontaneous R wave amplitude was 7.89 +/- 4.89 mV. The polarization signal (PS) was 3.8 +/- 3.04 mV, and evoked response (ER) was 8.15 +/- 4.57 mV at the predischarge testing. Follow-up telemetry was done at months 1, 3, 6, 12, and 18. The follow-up duration was 9.4 +/- 5 months (range 1-18 months). If the results of PS and ER measurements were acceptable for autocapture, it turned on at the 1-month visit. In six (40%) patients the results were found acceptable for autocapture function. Age, lead impedance, pacing threshold, intrinsic R wave measurement, lead age, fixation mechanism, and ER measurements were not statistically different in Autocapture suitable and not suitable groups. The main reason not to activate Autocapture had been increased PS. Any significant fluctuations were not observed in pacing threshold, lead impedance, ER, and PS during follow-up. In conclusion, previously implanted pacing leads may be compatible with the Autocapture function.
自动夺获功能可控制并优化起搏脉冲的幅度,从而节省能量。制造商建议使用一种特殊的低极化、低阈值双极百多力导线来实现自动夺获功能。本研究的目的是评估自动夺获功能与先前植入的起搏导线的兼容性。该研究纳入了15例需要更换脉冲发生器的患者(平均年龄13.6±3.4岁),他们接受了具有自动夺获功能的VVIR起搏器Regency SR+或DDDR起搏器Affinity DR或Integrity DR。新的脉冲发生器与先前植入的心室导线相连。植入时,起搏阈值在0.5毫秒时为1.0±0.35伏,导线阻抗为580±80欧姆,自发R波幅度为7.89±4.89毫伏。在放电前测试时,极化信号(PS)为3.8±3.04毫伏,诱发反应(ER)为8.15±4.57毫伏。在第1、3、6、12和18个月进行随访遥测。随访时间为9.4±5个月(范围1 - 18个月)。如果PS和ER测量结果对于自动夺获功能可接受,则在1个月随访时开启该功能。在6例(40%)患者中,发现自动夺获功能的结果可接受。在自动夺获功能适用组和不适用组中,年龄、导线阻抗、起搏阈值、固有R波测量、导线使用年限、固定机制和ER测量在统计学上无差异。未激活自动夺获功能的主要原因是PS升高。随访期间,起搏阈值、导线阻抗、ER和PS未观察到任何显著波动。总之,先前植入的起搏导线可能与自动夺获功能兼容。