To compare the effect on image quality of a 1.0 mol/L gadolinium chelate with that of two 0.5 mol/L gadolinium compounds.
Five healthy volunteers underwent a mono-station 3D MRA exam (Siemens SONATA, Erlangen, Germany) four times using four separate gadolinium preparations. All subjects first received a fixed volume of undiluted gadobutrol (1 mol/L), which corresponded to a dose between 0.1 and 0.15 mmol/kg body weight. This gadobutrol dosage was then diluted with saline into twice the volume and administered as a bolus at twice the injection rate. For Gd-DTPA and Gd BOPTA, because these contrast agents are 0.5 mol/L preparations, the volume and flow rate were doubled to match diluted gadobutrol volume and concentration. Quantitative and qualitative analysis of the angiographic data sets was performed on nine arterial segments.
Image quality was rated diagnostic for all image data sets without statistically significant differences between any of the compounds (P > 0.3). Quantitative measurements of Gd BOPTA (SNR: 81.15; CNR: 68.91) and both standard and diluted forms of gadobutrol (SNR: 84.33; CNR: 71.62; SNR(diluted): 79,23; CNR(diluted): 66.26) yielded significantly higher results (P < 0.02) in comparison with Gd-DTPA (SNR: 49.55; CNR: 38.24). The difference between either form of gadobutrol and Gd BOPTA was not shown to be statistically significant (P > 0.3), whereas both the SNR and CNR of standard gadobutrol were significantly higher than diluted gadobutrol.
Gadobutrol- and Gd BOPTA-MRA exams lead to improved delineation of the pelvic arterial morphology compared with MRA exams performed with Gd-DTPA.
比较1.0摩尔/升钆螯合物与两种0.5摩尔/升钆化合物对图像质量的影响。
五名健康志愿者使用四种不同的钆制剂,在单站3D MRA检查(西门子SONATA,德国埃尔朗根)中进行了四次检查。所有受试者首先接受固定体积的未稀释钆布醇(1摩尔/升),其剂量相当于0.1至0.15毫摩尔/千克体重。然后将该钆布醇剂量用生理盐水稀释至两倍体积,并以两倍注射速率推注给药。对于钆喷酸葡胺和钆贝葡胺,由于这些造影剂是0.5摩尔/升制剂,因此将体积和流速加倍以匹配稀释后的钆布醇体积和浓度。对九个动脉节段的血管造影数据集进行了定量和定性分析。
所有图像数据集的图像质量均被评为可用于诊断,任何化合物之间均无统计学显著差异(P>0.3)。与钆喷酸葡胺(SNR:49.55;CNR:38.24)相比,钆贝葡胺(SNR:81.15;CNR:68.91)以及钆布醇的标准形式和稀释形式(SNR:84.33;CNR:71.62;SNR(稀释):79.23;CNR(稀释):66.26)的定量测量结果显著更高(P<0.02)。钆布醇的任何一种形式与钆贝葡胺之间的差异均无统计学显著性(P>0.3),而标准钆布醇的SNR和CNR均显著高于稀释后的钆布醇。
与使用钆喷酸葡胺进行的MRA检查相比,钆布醇和钆贝葡胺MRA检查能更好地显示盆腔动脉形态。