Suppr超能文献

Impact of the model C and Automatic Positioning System on gamma knife radiosurgery: an evaluation in vestibular schwannomas.

作者信息

Régis Jean, Hayashi Motohiro, Porcheron Denis, Delsanti Christine, Muracciole Xavier, Peragut Jean Claude

机构信息

Department of Stereotactic and Functional Neurosurgery, Timone University Hospital, Marseille, France.

出版信息

J Neurosurg. 2002 Dec;97(5 Suppl):588-91. doi: 10.3171/jns.2002.97.supplement.

Abstract

OBJECT

The technical advances associated with the model C gamma knife include a robotized system enabling automatic positioning of the stereotactic coordinates. The purpose of this study was to analyze the clinical impact of this technical modification.

METHODS

The authors studied a sample of patients with vestibular schwannoma (VS). This sample included three groups treated using gamma knife radiosurgery. Group I comprised 21 patients with VS treated just before the installation of the Automatic Positioning System (APS). Group II included patients in Group I with new dose plans created using the APS (in other words, simulated dose plans). Group III consisted of a control group of 20 patients matched for tumor grade with the previous group and treated recently with the APS. Treatment times were calculated after correcting the time for each shot according to the age of the sources after reloading. The treatment times, including total time, irradiation time, and duration of the neurosurgical procedure, were analyzed. In addition, dose planning including number of isocenters, number of different collimators, malfunctions, and the conformity and selectivity indices were recorded. The trend was to reduce the mean number of collimator runs from 7.9 to 1.2 and to increase the mean number of shots from 7.9 to 15.6, mostly by using the 4-mm collimator exclusively. The APS-related conformity and selectivity were improved from 95 to 97% and from 78 to 84%, respectively. The total treatment time was reduced by 53%, and time required to interact with the patient in the room was considerably reduced (75%), giving the neurosurgeon greater freedom to perform other tasks during the treatment period. The reduction of the time spent by the neurosurgeon at work in the room was 84%. The total radiation time was increased by 54%.

CONCLUSIONS

The preliminary results of this study indicate that the robotization of the gamma knife is likely a major advance in radiosurgery.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验